Reliable Selection of Auditory Steady-State Responses Parameters under Bone,
Insertphone and Speaker Stimulations |
Dong-Geun Han1, Dukhwan Lim2 |
1Graduate School, Hallym Univeristy, Chuncheon, Korea 2Division of Speech Pathology & Audiology, Research Institute of Audiology & Speech Pathology, Hallym University, Chuncheon, Korea |
Correspondence |
Dukhwan Lim ,Tel: (033) 248-2217, Fax: (02) 6280-9133, Email: dlim@hallym.ac.kr
|
|
Received: October 21, 2005; Accepted: November 30, 2005. Published online: December 31, 2005. |
|
|
ABSTRACT |
The purpose of this study was to investigate the reliable selection of stimulus parameters in auditory steady-state responses (ASSR)
against electrical artifacts under clinical settings of air, bone and field stimulations. Data with these different transducers were obtained
from the mannequin showing the same electrical impedances as in the clinical recording sessions. The amplitude modulation was
fixed as 100% with the varied modulating frequency from 20 to 200 Hz and with the varied frequency modulation from 0 to 15%.
The results showed what points should be reviewed carefully in clinical recording sessions of ASSR data. First, the insert phone
stimulation led to the highest reliable range of ASSR stimulus parameters while the bone conduction stimulation exhibited the least
reliable ranges of stimulation parameters. Second, in this study, even insert phone stimulation led to some artifacts at the limited
frequencies of 1, 2 kHz with modulating frequency of 120 Hz. Thus, it would be necessary to pay special attention to reduce electrical
noises or to avoid these ranges to improve reliability. Third, the bone stimulation resulted in significant artifacts in the overall ranges
of stimulation parameters, so data from bone conduction ASSR suffered from the highest level of influences from electrical artifacts
and led to the least reliable interpretation. Fourth, the field stimulation exhibited some false positive responses at 4, 8 kHz with the modulating
frequencies of 120 and 160 Hz. Therefore, these ranges should be carefully considered with possible contamination from
electrical sources. Finally, the data showed the clinically important ranges against genuine electrical noises while the previous studies
contained responses with mixed noise sources such as electrical, vestibular, and other physiological origins. In conclusion, the clinically
safe and reliable ranges of stimulus parameters were estimated for all three stimulation modes. These data will be essential in
interpreting correct ASSR results and in designing more robust data aquisition strategies against various types of artifacts. |
Key Words:
Auditory steady-state response (ASSR).False positive.electrical noises. |